PSB-IC-SYS-1018 - AS-AP Autosampler Syringe Valve errors
Instances of Syringe Valve errors in AS-AP autosamplers before S/N 12040236 occur due to narrow encoder slots misaligning with sensors. The slots have been widened in new syringe pump assemblies since April 17, 2012, eliminating the error. If errors persist, replace the syringe pump.
Description:
Instances of Syringe Valve errors have occurred in the field. The cause can be due to the positioning of the valve in the center position to the flush bottle port.
Environment:
- AS-AP Autosampler Syringe Valve
- SN Prior to 12040236