PSB-IC-SYS-1029 - AS-AP Autosampler with Cooling Tray

Thermo Fisher Scientific provides a procedure to refill coolant and replace the reservoir cap in thermostatted AS-AP autosamplers. If cooling tray leakage is detected, follow the steps to refill the coolant reservoir or replace the carousel assembly. Contact Ion Chromatography Technical Support if problems persist.

Description: 

A field service procedure to refill the cooling solution of thermostatted AS-AP autosamplers and a field replacement procedure for the upgraded reservoir cap is provided in this PSB.

Environment:

  • AS-AP Autosampler with Cooling Tray
  • All SN