PSB-IC-SYS-1054R- AS-AP XZ ribbon cable breakage

A new motor controller cable assembly (PN: 60-074679) is being used in AS-AP autosamplers to connect the XZ Drive, Carousel, and Syringe Drive to the Motor Controller PCB. The new cable is more resistant to breakage and lasts longer. Replace damaged cables using the new version, especially in cases of XZ-arm related failures.

Description: 

AS-AP Autosampler as the X drive moves left and right, the ribbon cable bends at the top of the enclosure. This repetitive motion causes the wiring inside the cable to break, resulting in intermittent position errors.

Ribbon cable is now black, P/N 60-074679 (CBL ASSY,MTR CNTLR PWR/SIG,AS-AP).

Environment:

AS-AP Autosampler serial numbers before 15111335