PSB-IC-SYS-1069 - AS-AP New version of the AS-AP needle holder in the XZ drive assembly

A new AS-AP needle holder in the XZ drive assembly was implemented in October 2016, changing the NeedleForceAdjustmentDepth value from 3800 to 2784. Units shipped between October 2016 and January 2017 need this value updated to prevent needle port seal leakage. Future replacements of pre-October 2016 units also require this adjustment.

Description: 

A new version of the AS-AP needle holder in the XZ drive assembly has been implemented since 
October 2016.  With this change, the default value 3800 in the property for NeedleForceAdjustmentDepth has been updated to 2784, and this value is stored in EEPROM of the AS-AP main board.  

(Published February 2017)

 

Environment:

AS-AP Autosampler serial numbers before 17010623