AS-AP Firmware v 1.9.1 and Release Notes

Firmware version 1.9.1 for the AS-AP Autosampler was released in May 2024, superseding all previous versions. This update improves the starting height for alignment, raising the needle depth by 3.0 mm to minimize the risk of needle damage. It is recommended to update existing AS-AP Autosamplers to this new firmware for safer alignment.

Description: 

A new version of AS-AP firmware (version 1.9.1) has been released to address this issue and is included with all AS-AP instruments from serial number 2405305022 onward. Compared to the 1.9.0 firmware, the initial height of the 1.9.1 and Alignment of positions 1 and 2 has been increased by 3.0 mm when aligning the following:

·        Move to Align 1

·        Move to Align 2

Environment:

AS-AP Autosampler

 

 