60-065335 Rev B - Dionex AS-DV Autosampler - Preventive Maintenance Procedure
The Dionex AS-DV Autosampler Preventive Maintenance Procedure, revised in July 2016, outlines annual maintenance tasks. These include replacing the sampling tip and transfer line, lubricating the sampling mechanism if needed, and rebuilding the high-pressure valve. The procedure requires specific parts and tools, with a completion time of 30 to 60 minutes. Detailed steps ensure correct replacement and lubrication, maintaining system performance.
Description:
These instructions describe a preventive maintenance procedure for the Thermo Scientific Dionex AS-DV Autosampler. The procedure consists of replacing the sampling tip and transfer line the tubing between the sampling tip and the injection valve or optional high-pressure valve. Thermo Fisher Scientific recommends performing this procedure annually.
Environment:
- Dionex AS-DV Autosampler
Attachment(s):
| File | Last Modified |
|---|---|
| 60-065335 Rev B - Dionex AS-DV Autosampler - Preventive Maintenance Procedure.pdf | January 07, 2022 |