PSB-ICSP-1095 - Rev B - Integrion CM 7.2.10.23925 Windows 10 Communication Disconnects

Integrion HPIC systems may experience communication disconnects with Chromeleon 7.2.10.23925 on Windows 10 Dell computers, causing sequence interruptions. Users should avoid this Chromeleon version, remove Dell SupportAssist, or use a powered external USB hub or a non-Dell computer. Contact IC/SP PSE Group if issues persist.

Description: 

Dionex Integrion HPICs having communication disconnects with Chromeleon 7.2.10.23925 on Windows 10 Operating Systems in Dell computers

 

Environment:

Integrion HPIC and Chromeleon 7.2.10.23925