PSB-IC-SYS-1047 - ICS5000+ - DC-Lost Communication With Chromeleon
Thermo Fisher Scientific has identified a communication issue between the ICS-5000+ DC and Dell Optiplex 9020 PC when the DC upper compartment door is closed and temperature is on. The error can be resolved by using an externally powered USB 2.0 hub. Users in North America can order P/N 060392 or use any quality USB 2.0 hub with external power.
Description:
When a Thermo Fisher Scientific Dionex ICS-5000+ DC is connected with a Dell Optiplex 9020 PC, it is possible that the DC can lose communication with Chromeleon while the DC upper compartment door is closed and temperature is on.
Environment:
- Thermo Fisher Scientific Dionex ICS-5000+
- Dell Optiplex 9020 PC
- Chromeleon