065404 - Rev 04 - ICS-Modular System Installation Qualification

The ICS-Modular System Installation Qualification document outlines the procedures for verifying the proper installation of Dionex ICS-6000 and earlier systems. It includes steps for unpacking, placing modules, installing software, making electrical and USB connections, and setting up liquid lines. The document also describes the general system function tests for conductivity, absorbance, and electrochemical detectors. Example chromatograms are provided for reference.
The ICS-Modular System Installation Qualification document outlines the procedures for verifying the proper installation of Dionex ICS-6000 and earlier systems. It includes steps for unpacking, placing modules, installing software, making electrical and USB connections, and setting up liquid lines. The document also describes the general system function tests for conductivity, absorbance, and electrochemical detectors. Example chromatograms are provided for reference.

Description: 

This manual provides the installation instructions for the ICS-Modular System.

Environment:

Dionex ICS-6000 and these earlier modular systems:

  • Dionex ICS-3000
  • Dionex ICS-5000