How to replace the suppressor cable set for an ICS-6000 DC module
To replace the suppressor cable set for the ICS-6000 DC module, turn off and unplug the module, disconnect tubing, and move it to a spacious workbench. Disassemble the front door/frame, top cover, and various components to access the suppressor cable. Remove the old cable and install the new one by reversing the disassembly steps.
Description:
This article introduced the instructions of replacing the suppressor cable set for an ICS-6000 DC module when the connector or cable is broken, rusted or damaged.
Note: The DC suppressor cable set is not orderable. FSE who needs it please send the request to ICSP.ProductSupport@thermofisher.com.
Environment:
ICS-6000 DC
Attachment(s):
| File | Last Modified |
|---|---|
| Replacing the suppressor cable set for ICS6000 DC.pdf | April 11, 2024 |