PSB-ICSP-1068 - One-Wire Suppressor

The 1-Wire Enabled Suppressors use repurposed wires for memory tracking, causing incompatibility with some legacy Dionex IC systems, triggering an "SRS Alarm." Solutions include upgrading the IC system, using a Dionex RFC-10/30 controller or SCC-10, or using legacy Dionex SRS 300 suppressors.

Description: 

During initial development of the electrolytic suppressor, two wires in the cable were dedicated for connection to a thermistor to monitor suppressor temperature. In the mid 90’s the thermistor was removed from the suppressor leaving the two wires unused. In early 2016, these two wires were repurposed for use as a one-wire connection for memory tracking.

Environment:

Thermo Scientific Dionex DX-100 (Model DX 1-03) and DX-120

 

 