Preventive Maintenance for Integrion Ion Chromatography System

This page contains a collection of the parts, documents, tools and training needed to perform a Preventive Maintenance Service on an Integrion Ion Chromatography System.

Details

The article provides a list of the parts, procedures, factory documents, IES documents needed to perform a Preventative Maintenance or PM on a Dionex Integrion Ion Chromatography Systems.  This is also sometimes called Integrion HPIC system or Integrion Integrated IC system or PM Kit for Integrion or Integrion IC PM Kit.

Preventative Maintenance Kit or PM Kit Part Number

Dionex Integrion Preventative Maintenance Kit (part number 22153-62041)

Time to Complete Preventative Maintenance

4 hours

Documents Required

IES Preventative Maintenance Instructions:  ULS-DRS-PMI-0040 PMI ICSP Integrated IC ICS1X00 ICS2X00_Aquion_Integrion

IES Preventative Maintenance Checklist:  ULS-DRS-PMC-0159 PM Checklist ICSP  ICS1X00 ICS2X00 Aquion Integrion

Factory Preventative Maintenance Procedure:  22153-97004 Revision A March 22, 2016

Description of the work performed

During the Preventive Maintenance (PM) on the Dionex Integrion Ion Chromatography System, I followed the Preventive Maintenance Instructions (document number ULS-DRS-PMI-0040). The tasks performed included replacing O-rings, the end-line filter, pump check valve cartridges, priming valve seals, main seals, rebuilding the high-pressure  injection valve, and replacing the priming valve knob. System testing and calibration were conducted to ensure all parts were functioning correctly. All tasks were documented using the Preventive Maintenance Checklist (part number ULS-DRS-PMC-0159).

PM Kit Contents: of Dionex Integrion Preventative Maintenance Kit 22153-62041

Part Number Description Quantity
040695 O-ring

2

045987 End-line filter 1
045994 Pump check valve cartridge 2
062823 Next Due Date label 1
063382 Priming valve seal/seal wash seal 3
075768 Main Seal 2
075973 High-Pressure valve rebuild kit, 6-port 1
AAA-062056 Priming valve knob 1

Tools Required

Part Number Description
062336 1/2-in wrench
062338

3-mm hex key

079803 Syringe, 10 mL
042772 10-32 fitting plug

Preventative Maintenance Kits for Optional Auxiliary Valves

Part Number Description
074698 4-port auxiliary valve PM Kit
075973 6-port auxiliary valve PM Kit
079054 10-port auxiliary valve PM Kit