PSB-IC-SYS-1025 - ICS-Series Variable Wavelength Detector
Thermo Fisher Scientific announced a policy for ICS-Series Variable Wavelength Detectors (VWD) regarding mirror replacements. Mirrors degrade over time due to UV exposure, affecting performance. Detectors manufactured after March 2011 will receive free mirror kits, while older detectors will receive discounted replacements. The bulletin includes troubleshooting steps and recommendations for mirror replacement and filter cleaning, which should be performed by a certified service engineer.
Description:
Several Product Support Bulletins PSB SF 474, 474A, 561, and 593 have been issued by the HPLC product line in Germering recently which discuss optical elements on Variable Wavelength Detectors VWD; these specifically apply to the Ultimate HPLC series detectors. This PSB distills the information presented in those PSB releases and extends it to the ICS-Series VWD. The optics units of the ICSSeries VWD are identical to the Ultimate HPLC series VWD.
Environment:
- ICS-Series Variable Wavelength Detector
Attachment(s):
| File | Last Modified |
|---|---|
| PSB-IC-SYS-1025 - ICS-Series Variable Wavelength Detector.PDF | December 31, 2021 |