Skip to main content
Thermo Fisher Scientific

PSB-IC-SYS-1028 - ICS-Series Variable Wavelength Detector

Thermo Fisher Scientific addressed startup errors in ICS-Series Variable Wavelength Detectors caused by redesigned cooling fans. Errors such as "Lamp house cooling fan failure" and "System cooling fan failure" may appear. Upgrading the firmware to revision 1.25 and increasing fan startup voltages to 7V are recommended to resolve these issues. Ensure proper update procedures to avoid operational failures.

Description: 

Cooling fans used in the ICS-Series Variable Wavelength Detector VWD were recently redesigned. The time to power up the fans has become longer and sometimes the instrument’s internal diagnostics will produce an error message at startup. This PSB describes a combination of enhancements with a new cooling fan began shipping in Fall 2012 and firmware improvements.

Environment:

  • ICS-Series Variable Wavelength Detector

Affected SN:

  • Germering ID # 8066881 and above
  • Sunnyvale Serial # 13011185 and above

 

  • Was this article helpful?