PSB-IC-SYS-1036 - ICS-Series Variable Wavelength Detector
In July 2013, Thermo Fisher Scientific addressed an issue in ICS-Series Variable Wavelength Detectors (VWD) where some units fail at power-on with the error "ADC internal test failure - cycle power." The recommended recovery procedure involves disconnecting the USB cable, power cycling the detector, and waiting before reconnecting. If multiple attempts fail, replacing the main PCB or optical unit may be necessary. Specific part numbers for replacements are provided.
Description:
Some variable wavelength detectors will fail at power-on with a message ADC internal test failure - cycle power. This error is described in PSB-SF648, as related specifically to the Ultimate series VWDs. The error occurs in the same way for the ICS-Series VWD.
Environment:
- ICS-Series Variable Wavelength Detector
- 6074.3001 Main board (VWD- single channel) with extraction tool for flash memory
- 6074.1000 Optics VWD - single channel
- 6074.3010 Main board VWD - multiple channel with extraction tool for flash memory
- 6074.1010 Optics VWD - multiple channel
Attachment(s):
| File | Last Modified |
|---|---|
| PSB-IC-SYS-1036 - ICS-Series Variable Wavelength Detector.PDF | December 31, 2021 |

