PSB-SV-500 - ICS Series VWD Variable Wavelength Detector
In May 2008, Dionex addressed D2 lamp ignition failures in ICS Series Variable Wavelength Detectors (VWD). Symptoms include UV lamp ignition failure [LD] and Lamp voltage pre-set failed [LD] errors. Dionex is working on changes to the lamp power supply and firmware for a permanent fix. Failing lamps with less than 1000 hours will be exchanged on a goodwill basis. Firmware version 3.01 should be used to resolve voltage pre-set errors.
Description:
Handling of D2 lamp ignition failures of the ICS Series Variable Wavelength Detectors VWD as described in PSB SF327 also affect the ICS Series VWDs in the same manner. This PSB describes the current status and recommended actions to address D2 lamp ignition failures with ICS Series VWDs
Environment:
- ICS Series VWD Variable Wavelength Detector