SV-591R - ICS-Series VWD
Firmware revision 1.24 for ICS-Series Variable Wavelength Detectors (VWD) addresses lockup issues observed in previous versions. The update removes warning messages that appear in Chromeleon Audit Trail and improves stability, including fixing a bug that causes crashes after long periods of continuous operation. The firmware is compatible with all ICS-Series VWDs and Chromeleon versions. Upgrading to this firmware is recommended to resolve these issues.
Description:
ICS-Series Variable Wavelength Detectors may infrequently lock up and produce symptoms described in this PSB that are not documented in the CM audit trail.
Environment:
- ICS-Series VWD
- ICS-5000 Single Channel VWD 069116 10050150, 10120119, 10120312, 10120328, 10120404, 10120408
- ICS-5000 Four Channel VWD 069117 10120209
- ICS-5000 VWD 1 Channel 069116
- ICS-5000 VWD 4 Channel 069117
- ICS-3000 VWD 1 Channel 064377
- ICS-3000 VWD 4 Channel 065754
Attachment(s):
| File | Last Modified |
|---|---|
| SV 591R - ICS-Series VWD.PDF | January 01, 2022 |