ICS-Series VWD - Firmware 1.25 and Release Notes

Thermo Fisher Scientific released firmware rev 1.25 for ICS-Series Variable Wavelength Detectors to address prolonged power-up times due to changes in the cooling fan's control electronics. This update increases the detection grace period for fan failures, eliminating false error messages. The firmware is compatible with all ICS-3000/ICS-5000-VWD series devices and does not require requalification testing after the update. Field updates are recommended only if specific fan failure errors occur.

Description: 

The manufacturer of the VWD cooling fans has changed the fan’s control electronics. This results in a prolonged power up time.

Environment:

  • ICS-Series Variable Wavelength Detector Firmware

 