PSB-ICSP-1178-Inuvion Injection Valve and Door Plumbing Label Incorrect
Description:
There have been several customer reports from the Inuvion serial numbers documented above where the customer has complained of poor performance and/or poor chromatography when performing partial loop injections.
Investigations have found that there is an issue with the labeling of the injection valve and the door plumbing label. It is important to note that applications using full loop injections are not affected by this issue. However, it is recommended that the label is replaced on all the affected Inuvion modules to avoid any future issues.
Environment:
Inuvion Core
Inuvion
Inuvion RFIC
Recommendations:
The following steps should be taken for all Inuvion modules which have the incorrect label as below:
Inuvion Core S/N: 250251097 and lower
Inuvion and Inuvion RFIC S/N: 250251073 and lower
Replumb the Injection Valve:
Replumb the Injection Valve on all of the impacted Inuvions to ensure that the Injection Valve and the Door Label plumbing match. This involves checking that all connections are correct per the below correct label. In particular, switch the tubing from position W and S on any incorrect labelled Inuvion module, replacing any worn or damaged tubing, and ensuring that the system is free of leaks and blockages.
Replace the Injection Valve Label:
Replace the injection Valve label. Order 1 x P/N: B51004278 LABEL, INJECTION VALVE, INUVION per affected module. This involves removing the old label, affixing the new label in its place, and ensuring it is properly aligned and securely attached.
Replace the Inuvion system plumbing label on the inside of the door:
Replace the Inuvion system plumbing label on the inside of the door. This involves removing the old label, affixing the new label in its place, and ensuring that it is properly aligned and securely attached. Order 1 x correct label as per the Inuvion module listed below.
Inuvion Core: P/N B51000340 LABEL,PLUMBING,CORE,INUVION
Inuvion: P/N B51000341 LABEL,PLUMBING,INUVION
Inuvion RFIC: P/N B51000342 LABEL,PLUMBING,RFIC,INUVION