Installation of the Inuvion Eluent Monitoring Device

Provides instructions for the installation of the Inuvion eluent monitoring device. If the length of the USB cable is not long enough, see "Extending the Inuvion Eluent monitor USB cable".

Installation of the Inuvion Eluent Monitoring Device

The eluent monitoring device requires separate USB driver in order to be configured in Chromeleon:  https://www.silabs.com/developers/usb-to-uart-bridge-vcp-drivers?tab=downloads ,

  1. Select Windows drivers

clipboard_ea56213025be5d86e16d9b9735dd03b45.png

  1. Extract files and run x64bit exe file

clipboard_ee5c01c414f4d4c3b594c2937ebf00268.png

  1. Choose Eluent Monitor module in Chromeleon instrument configuration manager. Serial number should populate.
  2. Tare with empty bottle.
  3. Set threshold in Chromeleon.

NOTE: The same driver is also in the Packages section of Chromeleon 7.3.2 (and higher). Install the driver from installation media of Chromeleon 7.3.2 Packages "Silcon Labs USB to COM Port Driver CP210x" and run as administrator. Of course for the Inuvion it will also require 7.3.2Mub or higher.

NOTE: The Eluent Monitor should show green/red even if not configured in Chromeleon because of manufacturer calibrations (but will likely have higher threshold setting…when to show red)

Please also see this KA if the length of the USB cable is not long enough: Extending the Inuvion Eluent monitor USB cable - Thermo Fisher Scientific    

 

 

 