PSB-IC-SYS-1009R - IC Qualification Test Box II PN 22000-60001
Description:
An IC Qualification Test Box II PN 22000-60001 is available for the RFC-10, RFC-30, EG40, EG50, ICS-900, ICS-1000, ICS-1100, ICS-1500, ICS-1600, ICS-2000, ICS-2100, ICS-2500, ICS-3000, ICS-4000, ICS-5000, 5000+, Aquion and Integrion devices. This test box replaces the previous IC Validation/Test Box PN 72005.
*PLEASE NOTE: THE CURRENT FOR THE CR-TC TEST USING THE TEST BOX II (PURPLE BOX) SHOULD BE 18.4 mA (vs 96 mA on the older test box because of a change in the resistors used).
Environment:
- All SN
- IC Qualification Test Box II PN 22000-60001
- RFC-10
- RFC-30
- EG40
- EG50
- ICS-900
- ICS-1000
- ICS-1100
- ICS-1500
- ICS-1600
- ICS-2000
- ICS-2100
- ICS-2500
- ICS-3000
- ICS-4000
- ICS-5000
- ICS-5000+
- ICS-6000
- Aquion devices
- Integrion devices
Attachment(s):
| File | Last Modified |
|---|---|
| PSB-IC-SYS-1009R-OQPQ-ICQualificationTestBox-April2016.pdf | March 06, 2023 |