Skip to main content
Thermo Fisher Scientific

PSB-ICSP-1092 - Rev A - Dionex Pump Mechanisms Without Final Operational Qualification

Some Dionex pump mechanisms shipped between October 9-27, 2020, have incomplete FOQ paperwork and insufficient testing. Inspect arriving pump mechanisms for complete FOQ paperwork; if missing, mark as DoA and return for rework. Escalate issues to IC/SP PSE immediately.

Description: 

There have been instances of Dionex pump mechanisms shipping with incomplete Final Operational Qualification FOQ paperwork and insufficient testing. Further, it was discovered that in one instance, the complete FOQ was not performed on the pump mechanism prior to shipment, e.g., only one of four tests were performed and documented. Containment has been initiated at the factory starting on November 20th, 2020.

 

Environment:

  • ICS-6000
  • ICS-5000+
  • ICS-5000
  • Integrion