PSB-ICSP-1115 - Rev A - Integral SS-SP Main CPU PCB Not Programmed

Instances of unprogrammed main CPU boards for Integral SS and SP modules have been reported, indicated by connection issues or firmware version 0.0.0. These boards cannot be programmed in the field and must be returned for rework. Programmed boards will have a handwritten label starting with "D83XXX."

Description: 

There have been instances of main CPU boards used on the Integral SS and SP modules not being programmed. Reports of the SS or SP module not connecting or a firmware version of 0.0.0 are signs that the main CPU board is not programmed. Further, it is not possible to program a main CPU board in the field as it requires a test fixture and specialized software.

Environment:

Integral Main CPU PCBs for SS/SP