PSB-ICSP-1138 - Rev C - Bad CRC Aquion ICS-5000 + DC ICS-6000 DC

Description: 

The ICSP Product Support Engineering team has received multiple reports relating to an error message in the Chromeleon audit trail, “Error receiving data: Module communication error: bad CRC”. This occurs when there is a failure of the Cyclic Redundancy Check (CRC) which checks for communication between the instrument and Chromeleon CDS. This error has been reported for the Aquion, ICS-5000+, and the ICS-6000 and results in a USB communication disconnect between the instrument and Chromeleon software PC.

Investigation into the root cause of the CRC error is underway and an update to this PSB will be provided once the root cause and corrective action have been identified. In the interim, the below steps can be taken to prevent the CRC error from interrupting instrument operation.

 

Environment:

Aquion

ICS-5000+ DC

ICS-6000 DC

 

Recommendations:

To prevent the USB disconnect from occurring, the message error level can be changed in the Instrument configuration from abort to a warning or ignore as shown in PSB-ICSP-1138R-bad CRC Aquion_ICS-5000+ DC_ICS-6000 DC