PSB-ICSP-1142-Recall of Dionex Seven Anion Standard II Lot 22-158JKS

Thermo Fisher Scientific has issued a recall for Dionex Seven Anion Standard II, Lot 22-158JKS, due to incorrect Nitrate and Nitrite concentrations. Containment and recall actions were initiated on April 8, 2024, but some test boxes may have been shipped. Symptoms include low Nitrite peak response and high Nitrate peak response. If affected, contact your local order team for warranty replacement and further instructions. Subsidiaries/distributors should contact the IC/SP PSE Group.

Description: 

Following reports of problems with P/N: 057590 Dionex Seven Anion Standard II, Lot 22-158JKS manufacturing has determined that it contains incorrect Nitrate and Nitrite concentrations.  Containment has been initiated at the factory to recall this specific lot since April 08, 2024.

Environment:

P/N: 057590

Ion Chromatography

 

 