Thermo Fisher Scientific

PSB-ICSP-1160-Pump Waste Tubing Kit for ICS-6000 Pump and Integrion

Description: 

There have been reports of pump seal leaks on the Integrion IC system and the ICS-6000 SP and DP pump modules. 

 

The piston pump is a dynamic system in which the piston continuously moves through the pump seal. Surface tension and a wet layer on the piston surface may transfer into the seal wash area. This physical effect is inevitable and varies depending on system pressure; however, the rate of liquid transfer is extremely minimal and does not impact performance of the Instrument. Over extended runtime this liquid does accumulate and potentially overflow, if the Seal Wash Inlets and Outlets are not connected.  

 

·         Seeing fluid movement in the waste tubing or seal wash system, even when the seal wash pump is not in operation, is not an indication of a defective pump head or seal.

·         Abnormal chromatography (e.g., retention time shifts) or system pressure fluctuations may indicate that the system requires troubleshooting, which may include seal replacement or pump head maintenance.

 

However, if liquid within the seal wash cavity is not properly contained or routed to waste through tubing from the seal wash ports, it can accumulate and emerge as visible droplets from the ports. These droplets may trigger the leak sensor after dripping off the pump head.

 

This PSB is for users that no longer have their seal wash tubing for the ICS-6000 pumps or that are not using their seal wash.  It is also for Integrion systems that do not have the optional seal wash system installed.

 

To address this issue, the Pump Waste Tubing Kit (P/N B51006291) has been developed. This kit allows for the collection and routing of liquid from both seal wash outlets of the pump head to waste.

 

 

Environment:

ICS-6000

Integrion