PSB-ICSP-1162 Incorrect Sample Loop in Inuvion Shipkit
Description:
Due to a recent production error, the 25μl sample loop (P/N: 042857) was incorrectly manufactured using red PEEK tubing (0.005” ID) instead of the correct orange PEEK tubing (P/N: 042855, TUBING,.020X.062 PK,ORN). This issue may affect a limited number of Inuvion units.
Environment:
Inuvion
S/N Range: 250151037-250151087
Attachment(s):
| File | Last Modified |
|---|---|
| PSB-ICSP-1162 Incorrect Sample Loop in Inuvion Shipkit.pdf | March 07, 2025 |