PSB-ICSP-1166 - Rev A - Bad CRC ICS-5000+ ICS-6000 DC CM7-3-2MUe
Description:
Multiple reports have been received relating to an error message in the Chromeleon audit trail, “Error receiving data: Module communication error: bad CRC”. This error occurs due to a failure of the Cyclic Redundancy Check (CRC), which checks communication between the instrument and Chromeleon software PC. It has been reported for the ICS-5000+ DC and ICS-6000 DC modules, resulting in a USB communication disconnect error between the instrument and Chromeleon CDS software PC. Although issues related to the instrument PC settings and physical configuration have been identified in some cases, no resolution has been found for a high number of modules apart from the workaround documented in PSB-ICSP-1138 - Rev C - Bad CRC Aquion ICS-5000 + DC ICS-6000 DC.
Extensive analysis by factory software, firmware, electrical, and mechanical engineering teams has not yet isolated the root cause. Recognizing the impact on customers, a fix that automatically re-establishes communication has been implemented. This fix will be available in the upcoming releases of Chromeleon CDS 7.3.2 MUe and 7.2.10 MUj software. Please note that Chromeleon CDS 7.2.10 MUj is released June 2025 and documented in PSB-ICSP-1171 - Rev A - Bad CRC ICS-5000 ICS-6000 DC CM7-2-10MUj
The fix will be for the ICS-5000+ and ICS-6000 DC only at this time.
Environment:
ICS-5000+ DC
ICS-6000 DC
Recommendations:
It is recommended for customers currently running Chromeleon CDS 7.3.2 to upgrade the version to 7.3.2 MUe to avoid experiencing bad CRC disconnects. The Error level must be set to Abort in the Instrument configuration for the Error message “Error receiving data: %1” for the bad CRC fix to work correctly.
For Chromeleon CDS version 7.3.2, it is recommended to upgrade to version 7.3.2 MUe to avoid the bad CRC disconnect error.
Attachment(s):
| File | Last Modified |
|---|---|
| PSB-ICSP-1166-bad CRC ICS-5000+_ICS-6000_DC_CM7-3-2MUe.pdf | May 02, 2025 |

