PSB-ICSP-1182 New Pump CPU PCB for Integrion and ICS-6000

Description: 

Due to the EOL (End of life) of some critical electronic components in the PCB board level, a new pump CPU board (P/N: 22153-61028R) has been developed to replace both the ICS-6000 pump CPU PCB (P/N: 069824) and the Integrion Pump CPU PCB (P/N: 22153-61018R). The new pump CPU PCB performs equivalently to the previous pump CPU PCB.

Along with the new pump CPU PCB board, a new pump firmware 7.1.2 was also released to adapt the chip changes on the board. All ICS-6000 DP/SP pumps shipped after serial number 25080650 are equipped with the new firmware.
 

Chromeleon Compatibility: This new firmware version 7.1.2 will be fully compatible with Chromeleon version7.3.1 and later versions:
• CM 7.3.1 All versions.
• CM 7.3.2 All versions.
• CM 7.2.10 MUf and later versions.
• CM 7.4 All versions

Hardware Compatibility: This new firmware version 7.1.2 will be compatible with all ICS 6000, ICS 5000/5000+ and Integrion pump models.

Firmware file 7.1.2 can be downloaded in KB article: https://knowledge1.thermofisher.com/Chromatography/Ion_Chromatography/ICS-6000_HPIC_System/ICS-6000_Firmware_and_Release_Notes/ICS-6000_DPSP%2C_Integrion_Pump_Firmware_v7.1.2_Release_Notes 

Environment:

ICS-6000

Integrion

 

 