PSB-ICSP-1193-Rev A-Integrions RFICs with Chromeleon SE Licenses Running CM 7-2-10 MUj or Later
Description:
A report for an Integrion RFIC system that previously worked stopped allowing control after upgrading Chromeleon from version 7.2.10 MUa to MUj. The issue is due to certain older Chromeleon Single Edition (SE) licenses losing or lacking the required feature “ICS_2x00_GradientGeneration”, which is necessary to control Integrion RFIC systems.
Chromeleon SE licenses sold before November 2017 were intended only for non-RFIC systems. After the launch of Integrion RFIC, the SE license was redesigned into a limited Workstation Edition (WE) license that supports one Thermo Fisher instrument with the required features. As a result, upgrading to newer Chromeleon 7.2 versions may prevent older pre-November 2017 SE licenses from controlling Integrion RFIC systems.
License Clarification:
Some earlier Chromeleon Single Edition (SE) licenses may not include the feature ICS_2x00_GradientGeneration (ICS 2x00 Gradient Generation), which is required for control of Integrion RFIC systems. These licenses may have previously functioned with earlier Chromeleon 7.2.10 Maintenance Updates but can lose effective control capability following an upgrade to Chromeleon 7.2.10 MUj or later, where stricter license feature validation is enforced.
Earlier Chromeleon SE licenses issued or sold before the end of November 2017 only allowed control of non RFIC systems.
Environment:
Integrion RFICs
Chromeleon SE licenses
Chromeleon 7.2.10 MUj or later