PSB-ICSP-1194 Non-Thermo Scientific Combustion Remote Inject Signal Ignored in Integrion

Description: 

Customers who have an Integrion with a non-Thermo Scientific combustion continue to report samples being skipped intermittently during their analysis. This issue has only been observed after upgrading Chromeleon software to version 7.3.2 MUc or later, or version 7.2.10 MUi or later, as documented in PSB-ICSP-1151 (Integrion Disconnect CM 7.3.2 MUc) and PSB-ICSP-1153 (Integrion Disconnect CM 7.2.10 MUi).

The previous PSB documents were intended to address the Integrion “U0000 CD Data Timeout” disconnect issue by rapidly restoring Chromeleon communication with the Integrion system after the error is detected. However, during the investigation of the sample-skipping issue, it was observed that handling of the incoming TTL remote start trigger signal may be affected when both actions occur simultaneously.

The R&D software team investigated developing a permanent fix. However, these attempts were unsuccessfully.

As such, an alternative workaround has been developed to prevent sample skipping. This workaround utilizes an A2D adapter to handle incoming remote start trigger signals from non-Thermo Scientific combustion devices.

Additionally, field feedback suggests that increasing the IC START TTL signal duration to 3 seconds or longer in the combustion device may be effective. Implementing this change will require on-site support from the combustion device supplier’s technicians.

Environment:

Integrion

Chromeleon 7.3.2 MUc and later

Chromeleon 7.2.10 MUi and later

 

 