PSB-ICSP-1206 ICS-6000 Pump Seal Wash Not Working

Description: 

There have been several field reports that the seal wash system on new ICS-6000 DP/SP pumps does not operate as expected. The seal wash system runs once after power-up but does not operate thereafter.

The factory has confirmed that this issue is caused by a bug in firmware version 7.1.2. Firmware version 7.2.0 has been released to correct the issue.

Environment:

ICS-6000 Pumps

Firmware 7.1.2

 

 