Chromeleon peak shows n.a. amount results resolved by changing origin in calibration type
Customers report a peak or peaks shown n.a. in the amount column, their calibration curve type is set to 'ignore origin'. After changing Calibration Type origin the integration problem is resolved.
Issue
One or more selected peaks display "n.a." for the amount calculated even when the peak is labeled correctly and integrated. Other peaks also have their value calculated using the current processing method calibration.
Environment
Chromeleon Software
Chromeleon 7 Software
Resolution
Calibration Type was updated from "Ignore Origin" to "Force Through Origin". Values should now be visibile for peaks that were missing calibration.