Collecting the Pressure Trace of an AXP Configured in Chromeleon

Setting up a virtual channel to acquire AXP pressure trace

Issue

 Collecting the pressure trace of an AXP that is configured in Chromeleon

Environment

AXP

AXP-MS

Resolution

In order to collect the pressure trace of an AXP that is configured in Chromeleon, it is necessary to set up a Virtual Channel and enter the appropriate commands in the script of the Instrument Method. In the snip below, the formula "AXP.Pressure" was used because the device name in the instrument configuration was changed to AXP (from Pump) to provide clarity in the epanels, distinguishing the AXP from the analytical pump(s).

clipboard_e953f9466fbd8bc0326480d314ac2016e.png

virtual channel.jpg

 

 

 