Connecting AXP pumps to Chromeleon in stable COM port numbers

Issue

Using micro USB cables to connect the AXP pumps, it is observed that the built-in USB to RS232 converters in AXP pumps often receive different COM port numbers from Windows Operating System after a computer power cycle. This leads to incorrect assignments in Chromeleon, resulting in pumps receiving incorrect flow commands, which can damage suppressors and cause unreliable operation

Environment

AXP pumps

AXP-MS pumps

Chromeleon 

Resolution

There are two potential solutions:

  1. Use the serial command 'INx' to set/read the pump number, assigning different numbers to each of the three pumps. This will require changes in Chromeleon software.
  2. Use the RS232 port (RJ11 connector) on the pump and a USB-RS232 converter on the computer side. This method provides stable port numbers as the USB converters are independent of the pumps

AXP communication port.png

Analysis

RS232 port(RJ11 connector) on the pump provides stable COM port number than the Micro-USB port on the pump.

Cause

The OEM supplier of the AXP pump has observed the similar issue. The R&D team believes it is caused by the USB devices on the pump control board lacking unique IDs, leading Windows to identify them based on power-up order. Any slight timing differences can change the port number assignment.