Connecting AXP pumps to Chromeleon in stable COM port numbers
Issue
Using micro USB cables to connect the AXP pumps, it is observed that the built-in USB to RS232 converters in AXP pumps often receive different COM port numbers from Windows Operating System after a computer power cycle. This leads to incorrect assignments in Chromeleon, resulting in pumps receiving incorrect flow commands, which can damage suppressors and cause unreliable operation
Environment
AXP pumps
AXP-MS pumps
Chromeleon
Resolution
There are two potential solutions:
- Use the serial command 'INx' to set/read the pump number, assigning different numbers to each of the three pumps. This will require changes in Chromeleon software.
- Please refer to the KA article Flow Rate and Pressure Transducer Calibration Procedures for AXP pumps to enter Service mode then send commands to AXP pumps.
- Use the RS232 port (RJ11 connector) on the pump and a USB-RS232 converter on the computer side. This method provides stable port numbers as the USB converters are independent of the pumps

Analysis
RS232 port(RJ11 connector) on the pump provides stable COM port number than the Micro-USB port on the pump.
Cause
The OEM supplier of the AXP pump has observed the similar issue. The R&D team believes it is caused by the USB devices on the pump control board lacking unique IDs, leading Windows to identify them based on power-up order. Any slight timing differences can change the port number assignment.