Connecting the power cable to the motor PCB in the ICS-6000/Integrion pump blocks
Issue
Field reports indicate that FSEs have difficulty connecting the power cable to new replacement pump motor mechanisms in Integrion and ICS-6000 systems.
Some FSEs observed that the power connector color on the motor control PCB is white, whereas it was black on previous ICS-3000/5000 pumps.
Environment
ICS-6000 DP/SP Pump Module
Integrion Pump Drive
Resolution
The pump motor control PCB (P/N: 60-074622 ASSY, PCB, BLDC, DP, ICS-5000) has remained unchanged since its release to the San Jose manufacturing factory in June 2013. The supplier may have used different non-critical materials for production, but these changes do not require factory validation as they do not affect overall performance.

The production factory for ICS-6000 pump modules and mechanisms confirmed experiencing similar difficulties in connecting the pump motor power cable to the pump mechanism board. However, by slightly prying the pins upwards, the connector can still be plugged in. The white foam serves as a protective cushion to ensure the two pins do not touch the board, and it was not soldered onto the pump motor board

The difficulty FSE experienced with fitting the cable connector to the board connector is likely due to the small gap between the two pins and the green board, which makes it challenging to fit the red connector. The convex area circled in yellow obstructs the red connector when sliding it towards the pins, leading some engineers to believe the power connector cannot connect to the pins on the new board.

Factory technicians used tools like a flathead screwdriver or pliers to slightly enlarge the gap or raise the pin tips, allowing the power cable connector to fit the white connector on the board
Another option is to trim the convex to create an acute angle, reducing the obstruction at the front end.

Analysis
Engineers are permitted to use tools to carefully raise the pins when connecting the connector to the board.
Cause
The gap between the two pins and the green board is too small to fit the red connector