EG Test Box Not Being Recognized or Does Not Function
Issue
IC OQ/PQ Test box or IC Validation Test Box II is not being recognized or functioning when it is plugged in for testing.
Environment
Ion Chromatography
Resolution

Step 1: Inspect all electronic connections on the IC Test Box and on the IC system. Pay special attention to ensure all pins are present and are straight and not bent. A missing or bent pin can cause problems with recognition and proper operation.

Step 2: Ensure that a powered on digital multimeter is connected OR the jumper is in place. Note that this should be one or the other connection option. In no situation should both the Jumper and the multimeter be connected to the IC Test Box at the same time.

Step 3: Try testing the IC Test box by connecting to an EGC cartridge or IC system EGC connector.
Step 4: Connect the column outlet to the suppressor eluent inlet and flush the suppressor eluent outlet directly to waste for 20 min.
Step 5: Replace IC Test Box
Analysis
Maintenance and troubleshooting equipment required:
| Quantity | Part Number | Description |
| 1 | N/A | Digital Multimeter |
| 1 | 22000-60001 | IC Test Box II |