EG Test Box Not Being Recognized or Does Not Function

Issue

IC OQ/PQ Test box or IC Validation Test Box II is not being recognized or functioning when it is plugged in for testing. 

Environment

Ion Chromatography

Resolution

IC EG Test Box Not Recognized - image 01.png

Step 1:  Inspect all electronic connections on the IC Test Box and on the IC system.  Pay special attention to ensure all pins are present and are straight and not bent.  A missing or bent pin can cause problems with recognition and proper operation.

IC EG Test Box Not Recognized - image 02.jpg

Step 2:  Ensure that a powered on digital multimeter is connected OR the jumper is in place.  Note that this should be one or the other connection option.  In no situation should both the Jumper and the multimeter be connected to the IC Test Box at the same time.

IC EG Test Box Not Recognized - image 03.jpg

Step 3:  Try testing the IC Test box by connecting to an EGC cartridge or IC system EGC connector.

Step 4:  Connect the column outlet to the suppressor eluent inlet and flush the suppressor eluent outlet directly to waste for 20 min.

Step 5:  Replace IC Test Box

Analysis

Maintenance and troubleshooting equipment required:

Quantity Part Number  Description 
1  N/A Digital Multimeter 
1  22000-60001  IC Test Box II 

 

 