Eluent Expired Error Message Received In Chromeleon Audit Trail

Issue

"Eluent Expired" error is received for suppressor in Chromeleon 7 audit trail when pump flow is turned on.

Environment

ICS-5000/5000+
ICS-6000
Chromeleon 7.2
Chromeleon 7.3

Resolution

1. Reset the EC for the suppressor using the "F8"/"Commands" in Chromeleon. 

2. Navigate to the DC category in "Commands" and verify the "EC Remaining Hours". 

3. If the "EC Remaining Hours" has a value of 0, select the "Reset EC" located under "Suppressor 1" or "Suppressor 2" depending on which suppressor is associated with the error in the audit trail.