Error Code list of Dionex Integrion HPIC Systems
Issue
Integrion Error Codes were not listed in Thermo Scientific Dionex Integrion HPIC Service Manual.
Environment
Dionex Integrion HPIC Systems
Resolution
| Error Code Number | Description | Error Category |
|---|---|---|
| 1 | ERROR_TRAY_LEAK_WET | tray leak errors |
| 2 | ERROR_TRAY_LEAK_DRY | tray leak errors |
| 3 | ERROR_LOAD_INJECT_VALVE | high pressure valve errors |
| 4 | ERROR_CLEARED_LOAD_INJECT_VALVE | high pressure valve errors |
| 5 | ERROR_LOAD_INJECT_VALVE_BUSY | high pressure valve errors |
| 6 | ERROR_HP_VALVE_2 | high pressure valve errors |
| 7 | ERROR_CLEARED_HP_VALVE_2 | high pressure valve errors |
| 8 | ERROR_HP_VALVE_2_BUSY | high pressure valve errors |
| 9 | ERROR_I2C_BUS_FREEZING | I2C BUS |
| 10 | ERROR_TABLET_USER_ABORT_CM_RUN | Tablet |
| 11 | ERROR_COLUMN_BLOWER_DISCONNECTED | Column Compartment |
| 12 | ERROR_PUMP_IS_UNLINKED | Pump |
| 14 | ERROR_SUPPRESSOR_OVER_VOLTAGE | suppressor errors |
| 15 | ERROR_SUPPRESSOR_OVER_CURRENT | suppressor errors |
| 16 | ERROR_SUPPRESSOR_OVER_POWER | suppressor errors |
| 17 | ERROR_SUPPRESSOR_STOP_FOR_FLOW_RATE | suppressor errors |
| 19 | ERROR_CLEARED_SUPPRESSOR | suppressor errors |
| 20 | ERROR_SUPPRESSOR_NOT_INSTALLED | suppressor errors |
| 22 | ERROR_CHANGE_SUPPRESSOR_TYPE | suppressor errors |
| 23 | ERROR_SUPPRESSOR_CARTRIDGE_DISCONNECTED | suppressor errors |
| 24 | ERROR_SUPPRESSOR_CARTRIDGE_RECONNECTED | suppressor errors |
| 25 | ERROR_WRONG_DEVICE_CONNECTED_TO_CONNECTOR_A | suppressor errors |
| 28 | ERROR_SUPPRESSOR_2_OVER_VOLTAGE | suppressor 2 errors |
| 29 | ERROR_SUPPRESSOR_2_OVER_CURRENT | suppressor 2 errors |
| 30 | ERROR_SUPPRESSOR_2_OVER_POWER | suppressor 2 errors |
| 31 | ERROR_SUPPRESSOR_2_STOP_FOR_FLOW_RATE | suppressor 2 errors |
| 33 | ERROR_CLEARED_SUPPRESSOR_2 | suppressor 2 errors |
| 34 | ERROR_SUPPRESSOR_2_NOT_INSTALLED | suppressor 2 errors |
| 36 | ERROR_CHANGE_SUPPRESSOR_TYPE_2 | suppressor 2 errors |
| 37 | ERROR_SUPPRESSOR_2_CARTRIDGE_DISCONNECTED | |
| 38 | ERROR_SUPPRESSOR_2_CARTRIDGE_RECONNECTED | |
| 42 | ERROR_DATA_COLLECTION_NOT_ENOUGH_DATA_POINTS | data collection errors |
| 43 | ERROR_INCOMPATIBLE_CD_DETECTOR | data collection errors |
| 49 | ERROR_ED_CELL_WORKING_ELECTRODE_DISCONNECTED | Electrochemical detector errors |
| 50 | ERROR_ED_CELL_WORKING_ELECTRODE_RECONNECTED | Electrochemical detector errors |
| 51 | ERROR_ED_CELL_REFERENCE_ELECTRODE_DISCONNECTED | Electrochemical detector errors |
| 52 | ERROR_ED_CELL_REFERENCE_ELECTRODE_RECONNECTED | Electrochemical detector errors |
| 53 | ERROR_ED_CELL_PH_OFFSET_CALIBRATION | Electrochemical detector errors |
| 54 | ERROR_ED_CELL_PH_SLOPE_CALIBRATION | Electrochemical detector errors |
| 55 | ERROR_ED_CELL_CURRENT_EXCEEDS_LIMITS | Electrochemical detector errors |
| 56 | ERROR_ED_CELL_CURRENT_EXCEEDS_LIMITS_CLEAR | Electrochemical detector errors |
| 57 | ERROR_ED_CELL_OFF_FOR_ZERO_FLOW | Electrochemical detector errors |
| 58 | ERROR_INCOMPATIBLE_ED_DETECTOR | Electrochemical detector errors |
| 64 | ERROR_QD_OFF_FOR_ZERO_FLOW | charge detector error |
| 65 | ERROR_QD_CELL_OPTION_DISCONNECTED | charge detector error |
| 66 | ERROR_QD_CELL_OPTION_RECONNECTED | charge detector error |
| 67 | ERROR_INVALID_QDC_SERIAL_NUMBER | charge detector error |
| 68 | ERROR_INCOMPATIBLE_QD_DETECTOR | charge detector error |
| 75 | ERROR_EGC_OVER_CURRENT | EG errors |
| 76 | ERROR_EGC_OVER_VOLTAGE | EG errors |
| 77 | ERROR_EGC_OVER_POWER | EG errors |
| 78 | ERROR_EGC_INVALID_CONCENTRATION | EG errors |
| 79 | ERROR_EGC_INVALID_FLOWRATE | EG errors |
| 80 | ERROR_EGC_CALIBRATION_ERROR | EG errors |
| 81 | ERROR_EGC_CARTRIDGE_DISCONNECTED | EG errors |
| 82 | ERROR_EGC_INVALID_SERIAL_NUMBER | EG errors |
| 83 | ERROR_EGC_ACTIVATION_DATE_NOT_VALID | EG errors |
| 84 | ERROR_EGC_ION_COUNT_NOT_VALID | EG errors |
| 85 | ERROR_EGC_WRONG_CARTRIDGE_INSTALLED | EG errors |
| 86 | ERROR_EGC_CARTRIDGE_RECONNECTED | EG errors |
| 87 | ERROR_EGC_ERROR_CLEAR | EG errors |
| 89 | ERROR_EG_1_WIRE_ACCESS_ERROR | EG errors |
| 90 | ERROR_EG2_1_WIRE_ACCESS_ERROR | EG errors |
| 91 | ERROR_EG3_1_WIRE_ACCESS_ERROR | EG errors |
| 94 | ERROR_EGC_2_OVER_CURRENT | EG2 errors |
| 95 | ERROR_EGC_2_OVER_VOLTAGE | EG2 errors |
| 96 | ERROR_EGC_2_OVER_POWER 96 | EG2 errors |
| 97 | ERROR_EGC_2_INVALID_CONCENTRATION | EG2 errors |
| 98 | ERROR_EGC_2_INVALID_FLOWRATE | EG2 errors |
| 99 | ERROR_EGC_2_CALIBRATION_ERROR | EG2 errors |
| 100 | ERROR_EGC_2_CARTRIDGE_DISCONNECTED | EG2 errors |
| 101 | ERROR_EGC_2_INVALID_SERIAL_NUMBER | EG2 errors |
| 102 | ERROR_EGC_2_ACTIVATION_DATE_NOT_VALID | EG2 errors |
| 103 | ERROR_EGC_2_ION_COUNT_NOT_VALID | EG2 errors |
| 104 | ERROR_EGC_2_WRONG_CARTRIDGE_INSTALLED | EG2 errors |
| 105 | ERROR_EGC_2_CARTRIDGE_RECONNECTED | EG2 errors |
| 106 | ERROR_EGC_2_ERROR_CLEAR | EG2 errors |
| 113 | ERROR_EGC_3_OVER_CURRENT | EG3 errors |
| 114 | ERROR_EGC_3_OVER_VOLTAGE | EG3 errors |
| 115 | ERROR_EGC_3_OVER_POWER | EG3 errors |
| 116 | ERROR_EGC_3_INVALID_CONCENTRATION | EG3 errors |
| 117 | ERROR_EGC_3_INVALID_FLOWRATE | EG3 errors |
| 118 | ERROR_EGC_3_CALIBRATION_ERROR | EG3 errors |
| 119 | ERROR_EGC_3_CARTRIDGE_DISCONNECTED | EG3 errors |
| 120 | ERROR_EGC_3_INVALID_SERIAL_NUMBER | EG3 errors |
| 121 | ERROR_EGC_3_ACTIVATION_DATE_NOT_VALID | EG3 errors |
| 122 | ERROR_EGC_3_ION_COUNT_NOT_VALID | EG3 errors |
| 123 | ERROR_EGC_3_WRONG_CARTRIDGE_INSTALLED | EG3 errors |
| 124 | ERROR_EGC_3_CARTRIDGE_RECONNECTED | EG3 errors |
| 125 | ERROR_EGC_3_ERROR_CLEAR | EG3 errors |
| 130 | ERROR_CRTC_1_INVALID_SERIAL_NUMBER | CR-TC errors |
| 131 | ERROR_CR_TC_STOP_FOR_ZERO_FLOW | CR-TC errors |
| 132 | ERROR_CR_TC_OVER_CURRENT | CR-TC errors |
| 133 | ERROR_CR_TC_SHORT_CIRCUIT | CR-TC errors |
| 140 | ERROR_CRTC_2_INVALID_SERIAL_NUMBER | CR-TC 2 errors |
| 141 | ERROR_CR_TC_2_STOP_FOR_ZERO_FLOW | CR-TC 2 errors |
| 142 | ERROR_CR_TC_2_OVER_CURRENT | CR-TC 2 errors |
| 143 | ERROR_CR_TC_2_SHORT_CIRCUIT | CR-TC 2 errors |
| 149 | ERROR_PS21_OVER_CURRENT | PS21 errors |
| 150 | ERROR_PS21_OVER_VOLTAGE | PS21 errors |
| 151 | ERROR_PS21_OVER_POWER | PS21 errors |
| 152 | ERROR_PS21_STOP_FOR_ZERO_FLOW | PS21 errors |
| 153 | ERROR_PS21_CONFIGURATION_CHANGE | PS21 errors |
| 159 | ERROR_PS22_OVER_CURRENT | PS22 errors |
| 160 | ERROR_PS22_OVER_VOLTAGE | PS22 errors |
| 161 | ERROR_PS22_OVER_POWER | PS22 errors |
| 162 | ERROR_PS22_STOP_FOR_ZERO_FLOW | PS22 errors |
| 163 | ERROR_PS22_CONFIGURATION_CHANGE | PS22 errors |
| 169 | ERROR_CRTC_EPM_1_1_WIRE_RESET_TO_DEFAULT | CRTC/EPM 1&2 one wire memory errors |
| 170 | ERROR_CRTC_EPM_1_CARTRIDGE_DISCONNECTED | CRTC/EPM 1&2 one wire memory errors |
| 171 | ERROR_CRTC_EPM_1_CARTRIDGE_RECONNECTED | CRTC/EPM 1&2 one wire memory errors |
| 172 | ERROR_CRTC_EPM_2_1_WIRE_RESET_TO_DEFAULT | CRTC/EPM 1&2 one wire memory errors |
| 173 | ERROR_CRTC_EPM_2_CARTRIDGE_DISCONNECTED | CRTC/EPM 1&2 one wire memory errors |
| 174 | ERROR_CRTC_EPM_2_CARTRIDGE_RECONNECTED | CRTC/EPM 1&2 one wire memory errors |
| 188 | ERROR_FLASH_MEMORY_RESET_TO_DEFAULTS | memory initialization errors |
| 189 | ERROR_CD_FRAM_RESET_TO_DEFAULTS | memory initialization errors |
| 190 | ERROR_ED_FRAM_RESET_TO_DEFAULTS | memory initialization errors |
| 191 | ERROR_QD_FRAM_RESET_TO_DEFAULTS | memory initialization errors |
| 192 | ERROR_QD_1_WIRE_RESET_TO_DEFAULT | memory initialization errors |
| 193 | ERROR_3_CHANNEL_BOARD_MEMORY_RESET_TO_DEFAULTS | memory initialization errors |
| 194 | ERROR_2_CHANNEL_BOARD_MEMORY_RESET_TO_DEFAULTS | memory initialization errors |
| 195 | ERROR_EG_1_WIRE_RESET_TO_DEFAULT | memory initialization errors |
| 196 | ERROR_EG_2_1_WIRE_RESET_TO_DEFAULT | memory initialization errors |
| 197 | ERROR_EG_3_1_WIRE_RESET_TO_DEFAULT | memory initialization errors |
| 198 | ERROR_COLUMN_HEATER_EEPROM_RESET_TO_DEFAULT | memory initialization errors |
| 199 | ERROR_TEC_EEPROM_RESET_TO_DEFAULT | memory initialization errors |
| 206 | ERROR_CELL_HEATER_EXCEEDS_SAFE_TEMPERATURE | cell heater errors |
| 207 | ERROR_CELL_TEMPERATURE_TOO_LOW | cell heater errors |
| 211 | ERROR_COMPARTMENT_PELTIER_EXCEEDS_SAFE_TEMPERATURE | Thermal compartment errors |
| 212 | ERROR_COMPARTMENT_PELTIER_CALIBRATION | Thermal compartment errors |
| 213 | ERROR_COMPARTMENT_PELTIER_TEMPERATURE_TOO_LOW | Thermal compartment errors |
| 214 | ERROR_COMPARTMENT_DOOR_OPENED | Thermal compartment errors |
| 215 | ERROR_COMPARTMENT_DOOR_CLOSED | Thermal compartment errors |
| 219 | ERROR_COLUMN_HEATER_EXCEEDS_SAFE_TEMPERATURE | column heater errors |
| 220 | ERROR_COLUMN_HEATER_CALIBRATION | column heater errors |
| 221 | ERROR_COLUMN_HEATER_TEMPERATURE_TOO_LOW | column heater errors |
| 222 | ERROR_COLUMN_COMPARTMENT_DOOR_OPENED | column heater errors |
| 223 | ERROR_COLUMN_COMPARTMENT_DOOR_CLOSED | column heater errors |
| 224 | ERROR_COLUMN_TEMPERATURE_SENSOR | column heater errors |
| 225 | ERROR_TEC_TEMPERATURE_SENSOR | column heater errors |
| 226 | ERROR_AMBIENT_TEMPERATURE_SENSOR | column heater errors |
| 227 | ERROR_DOOR_INITIATED_SCAN | RFID_SCAN errors |
| 228 | ERROR_RFID_PRODUCT_TYPE_NOT_RECOGNIZED | RFID_SCAN errors |
| 229 | ERROR_RFID_SCAN_FAILED | RFID_SCAN errors |
| 230 | ERROR_RFID_WRITE_FAILED | RFID_SCAN errors |
| 231 | ERROR_RFID_READ_FAILED | RFID_SCAN errors |
| 232 | ERROR_RFID_USER_DATA_ACCESS_FAILED | RFID_SCAN errors |
| 233 | ERROR_RFID_MEM_ACCESS_DISABLED | RFID_SCAN errors |
| 234 | ERROR_RFID_NOT_READY | RFID_SCAN errors |
| 235 | ERROR_RFID_TAG_1_CHECKSUM_FAILED | RFID_SCAN errors |
| 236 | ERROR_RFID_TAG_2_CHECKSUM_FAILED | RFID_SCAN errors |
| 237 | ERROR_RFID_TAG_3_CHECKSUM_FAILED | RFID_SCAN errors |
| 238 | ERROR_RFID_TAG_4_CHECKSUM_FAILED | RFID_SCAN errors |
| 239 | ERROR_RFID_TAG_5_CHECKSUM_FAILED | RFID_SCAN errors |
| 240 | ERROR_RFID_TAG_6_CHECKSUM_FAILED | RFID_SCAN errors |
| 241 | ERROR_RFID_TAG_7_CHECKSUM_FAILED | RFID_SCAN errors |
| 242 | ERROR_RFID_TAG_8_CHECKSUM_FAILED | RFID_SCAN errors |
| 243 | ERROR_RFID_TAG_9_CHECKSUM_FAILED | RFID_SCAN errors |
| 244 | ERROR_RFID_TAG_10_CHECKSUM_FAILED | RFID_SCAN errors |
| 245 | ERROR_RFID_WRITE_ABORT | RFID_SCAN errors |
| 246 | ERROR_RFID_READ_ABORT | RFID_SCAN errors |
| 247 | ERROR_RFID_SCAN_ABORT | RFID_SCAN errors |