How to Perform Preventive Maintenance on AXP Pumps

This article provides step-by-step instructions on how to perform preventive maintenance (PM) on AXP pumps, including the necessary part numbers and procedures to ensure optimal performance and longevity of the equipment.

Issue

Preventive maintenance (PM) is required to ensure the proper functioning and longevity of AXP pumps. This article outlines the steps to perform PM on AXP pumps, including the replacement of seals and check valves.

Environment

  • AXP Pumps
  • Part Number: 061041 (Seals)
  • Part Number: 061035 (Check Valves)

Resolution

  1. Identify the Parts Needed:
    • Seals: Part Number 061041
    • Check Valves: Part Number 061035 (if applicable)
  2. Preparation:
    • Ensure the pump is turned off and disconnected from the power source.
    • Gather necessary tools and replacement parts.
  3. Replace the Seals:
    • Locate the seals on the pump.
    • Remove the old seals carefully.
    • Install the new seals (PN 061041) in the correct position.
    • Ensure the seals are properly seated and secured.
  4. Check Valves (if applicable):
    • Although it is not typically required to replace the check valves during every PM, inspect them for any signs of wear or damage.
    • If replacement is needed, remove the old check valves (PN 061035).
    • Install the new check valves in the correct position.
    • Ensure the check valves are properly seated and secured.
  5. Final Checks:
    • Reconnect the pump to the power source.
    • Turn on the pump and check for any leaks or unusual noises.
    • Ensure the pump is operating smoothly and efficiently.
    • Run an example chromatogram (c-gram) for the customer application to verify proper functioning.

 

 