How to Perform a Suppressor Leak Check

How to Perform a Suppressor Leak Check

Issue

Users may need to determine whether an ion chromatography suppressor has developed an internal or external leak.  This article provides a simple method to verify suppressor integrity and identify if a suppressor has an internal leak.

Environment

  • Ion Chromatography systems
  • IC suppressors
  • Water or standard eluent supply

Resolution

  1. Pump water or eluent through the system.
  2. Begin pumping liquid through the suppressor.
  3. Remove the Regen In and Regen Out connections on the suppressor.
  4. Observe the suppressor and fluid flow paths carefully.
  5. After a period to allow flow to stabilize, verify that liquid exits only from the Eluent Out port of the suppressor. There should be no flow from the Regen ports or seams of the suppressor.
  6. If liquid leaks from the suppressor seams, backflushing the suppressor may correct the problem. 
  7. If liquid exits through the regeneration ports, the suppressor membrane has likely ruptured.
  8. Replace the suppressor if membrane damage is confirmed.
  9. Before installing a new suppressor, confirm that the backpressure on the suppressor is less than 200psi. See the start up  instructions in the suppressor operator manual, https://docs.thermofisher.com/r/NGES...9484939v2en-US 

Analysis

During internal troubleshooting discussions, it was noted that some users and support personnel may not be aware of the basic suppressor leak check process. Capturing this procedure in the knowledgebase improves accessibility and consistency for future troubleshooting activities.

Cause

A suppressor leak may be caused by one or more of the following:

  • Damage to suppressor seams
  • Internal membrane puncture or failure
  • Pressure-related damage
  • Chemical degradation over time