Inject Mode for AS-AP TLV Calibration should be PushFull

Issue

Calibrating the TLV on an AS-AP

Environment

 AS-AP

Resolution

For TLV calibration on an AS-AP, confirm the inject mode is set to PushFull (or PushSeqFull, if appropriate).  If it is set to PushPartial, exit the TLV Calibration screen. Click on the More Options button, and select PushFull. Restart the TLV calibration at Step 1. 