Installation Instructions for Dual Eluent Generator (EG) Degassers Setup for Electrochemical Detection

Issue

The eluent generation feature in Integrion and ICS-6000, coupled with pulsed amperometric detection (HPAE-PAD) is a widely used method.  For some applications, such as the separation of Fluorodeoxyglucose (FDG) from Fluorodeoxymannose (FDM), which has the Eluent Generator (EG) running constantly at 100mM and 1mL/min, not all residual gases are removed by the eluent degasser during the runs. As a result, some amount of residual gases dissolved in the eluent can cause an unwanted peak and baseline disturbances.

Environment

  • ICS-6000
  • Integrion
  • HPAE-PAD
  • Amperometric Detection
  • Electrochemical Detection 

Resolution

To eliminate a ghost peak and baseline disturbances, the following steps are required to efficiently remove residual gases form the RFIC-EG eluent setup:

  1. Degas di-H2O in the bottle thoroughly and pressurize the di-H2O bottle with inert gas (N2 or He) at 8 to 10 psi.
  2. Assemble and use a Dual Eluent Generator (EG) Degasser according to the attached installation instructions. Please note that the previous generation of EG degassers (P/N: 062137, Assy, Degas, EG, Analytical, ICS3) is used in this setup; thus the total backpressure should be below 3000 psi.

 

 