Part number and replacement steps for the ICS-6000 pressure transducer

This article provides instructions on replacing the pressure transducer and secondary pump head in ICS-6000 Pumps, including part numbers and descriptions for both Analytical and Capillary pump heads.

Issue

How to replace the pressure transducer for an ICS-6000 pump.

How to replace the secondary pump head for an ICS-6000 pump.

Note:  The pressure transducer is built into the secondary pump head assembly. 

What is the part number for the ICS-6000 Pump pressure transducer?

What is the part number for the ICS-6000 Pump secondary pump head?

What is the part number for the ICS-6000 Pump capillary secondary pump head?

Note: The mounting screws for the secondary pump head are unavailable for ordering and can be purchased at a local hardware store. The mounting screws are included with purchase of complete pump heads.

Environment

ICS-6000 SP Pump - Analytical and Capillary

ICS-6000 DP Pump - Analytical and Capillary

Part Number

The ICS-6000 Analytical pump head assembly part number is 083482 and the SAP description for this is:  "HEAD ASSY,SECONDARY,PMP,DP,SP "

The ICS-6000 Capillary pump head assembly part number is "075458" and the SAP description for this is: "ASSY,HEAD,SEC,HI PRES UPGRD,TESTED"

Resolution

The following steps are used to replace the Secondary Pump Head Assembly:

  1. Turn off the pump flow.
  2. Monitor the Current Pressure reading on the pump ePanel on the Chromeleon ePanel Set. When the system pressure reaches zero:
    • Disconnect the pump from Chromeleon by clicking "Connected" on the pump ePanel.
    • Press the POWER switch on the back of the DP/SP to turn off the pump .
  3. Open the DP/SP door to access the mechanical components .
  4. Remove the 2 mounting screws installed on the pump head.
  5. Wear clean room gloves to prevent contamination of pump parts while disassembling and reassembling the pump head.
  6. Disconnect all tubing connections to the pump head with the defective piston seal.
  7. Remove the pump head:
    • Be sure to disconnect the pressure transducer cable from position J203 on the main/CPU board for the top pump.
    • Be sure to disconnect the pressure transducer cable from position J202 on the main/CPU board for the bottom pump.
  8. Install the new pump head:
    • Place the new pump head back on the pump.
    • Reconnect transducer cable to original position on main/CPU board.
    • Reinstall the screws in the pump head using a 3.0 mm hex key (P/N 062338). Tighten the screws just until they come into contact with the pump head, then tighten another one-quarter to one-half turn, one-eighth of a turn at a time.
  9. Reconnect all tubing connections to the pump head. Tighten connections fingertight, then tighten an additional one-quarter turn only.
  10. Push the component mounting panel back into the enclosure using the handle in the center of the panel. Close the DP/SP door.
  11. Press the POWER button on the front of the DP/SP to turn on the pump.
  12. Reconnect the pump to Chromeleon by clicking "Connected" on the pump ePanel.
  13. Turn on the pump flow and verify pump pressure.