Part number for the AXP pump Inlet and Outlet filter elements
Issue
What is the part number for the AXP pump inlet filter elements kit?
What is the part number for the AXP pump outlet filter elements kit?
How do I prevent loss in flow through AXP pump filter elements?
Environment
- AXP
- AXP-MS
Part Number
The part number for an AXP/AXP-MS pump inlet filter elements kit is C0960-01-00106 and the SAP description for this is: "KIT,INLET,FILTER ELEMENTS, TELEDYNE, ROHS".
The part number for an AXP/AXP-MS pump outlet filter elements kit is C0960-01-00105 and the SAP description for this is: "KIT,OUTLET,FILTER ELEMENTS,TELEDYNE,ROHS".
Part Image
Resolution
Inlet filters should be checked periodically to ensure that they are clean and do not restrict flow. A restriction could cause cavitation and flow loss in the pump. Two problems that can plug an inlet filter are microbial growth and impure solvents. To prevent microbial growth, use at least 10% to 20% organic solvent in the mobile phase or add a growth-inhibiting compound. If you pump 100% water or an aqueous solution without any inhibitors, microbes will grow in the inlet filter over time, even if you make fresh solution every day. Always use well-filtered, HPLC-grade solvents for your mobile phase.