Request part number for a 1 ml sample loop, AS Autosamplers
Issue
How to create a 1 ml sample loop or order premade sample loops.
Environment
ICS system
Ion Chromatography (IC)
Injection
Injection valve
Injection loop
Resolution
1. Asked if part number 052305 ; tubing, PEEK, Green 0.003 i.d. 5 feet would make a 1 ml loop.
2. Checked attached chart and the customer would need 86.32 inches for a 1 ml loop so their part would not be suitable.
3. Gave customer part number 052682 for a 1000 microliter loop. Please see attachments.
Attachment(s)
| File | Last Modified |
|---|---|
| SPL LOOPS.pdf | May 04, 2022 |
| sample loops.PNG | May 04, 2022 |