Setting Up Simultaneous Injections Using ASDV with Two IC Systems
Issue
Setting up simultaneous injections using ASDV with two IC systems requires careful configuration to ensure accurate and efficient operation. Misconfigurations can lead to issues such as carryover.
Environment
- AS-DV
- Inuvion Systems
- ICS Systems
Resolution
- Instrument Method Configuration:
- Ensure that the instrument methods correspond to the correct channels to run both instrument methods at the same time.
- For example, Configure the anion method for Channel 1 and the cation method for Channel 2 to collect data from both channels simultaneously.
- Panel Control Setup:
- Integrate all systems into one set of e-panels (i.e. one instrument configuration) to manage the simultaneous injections effectively.
- Splitting the Sampler and transfer lines:
- Use the Simultaneous injection kit, 060258. Instructions are included with the kit, but can also be found here: AS-DV Simultaneous Injection Instructions
- Ensure the split is configured correctly to avoid cross-contamination and carryover issues.
- Series Configuration (alternate plumbing):
- An alternative method is to configure the systems in series. Direct the ASDV sample line to the Sample port on the first system and route the waste from the first system to the second system's sample port injection valve. This setup can help mitigate carryover issues.
- Documentation and Support:
- Refer to the official documentation for simultaneous injection setups. AS-DV Simultaneous Injection Instructions
- Contact technical support or experienced colleagues for additional guidance if needed.
Example Scenario
In a setup involving two Inuvion systems with an ASDV, configure the anion method for Channel 1 (Inuvion 1) and the cation method for Channel 2 (Inuvion 2). Ensure the sample is split correctly to both Inuvions. In cases of carryover issues, consider setting up the systems in series to direct the waste from the first system to the second system's injection valve.

