Triggering an ICS-6000 DP pump to stop the flow when the linked AXP pump motor stops by an error
This article describes how to trigger an ICS-6000 DP pump to stop the flow when the linked AXP/AXP-MS pump motor stops by error. In many situations, the AXP/AXP-MS pump acts as an external water delivery pump for an ICS-6000 system. To protect the electrolytic devices that are connected to the ICS-6000 pump, the analytical pump needs to be turned off to automatically shut the electrolytic devices off once the AXP pump stops delivering the regenerant solution (external water).
Issue
This article describes how to trigger an ICS-6000 DP pump to stop the flow when the linked AXP/AXP-MS pump motor stops by error.
In many situations, the AXP/AXP-MS pump acts as an external water delivery pump for an ICS-6000 system. To protect the electrolytic devices that are connected to the ICS-6000 pump, the analytical pump needs to be turned off to automatically shut the electrolytic devices off once the AXP pump stops delivering the regenerant solution (external water).
Top pump and bottom pump need to be triggered to stop separately in ICS-6000 dual pump models.
Resolution
- Configure the AXP/AXP-MS pump and ICS-6000 pump module in Chromeleon.
- Ensure three ICS-6000 pump TTL "Inputs" are NOT checked in the configuration.
- Locate the 25-pin D-Sub Port (Female) from the ICS-6000 Pump Shipkit.
- Locate the 10-in Terminal Board Connector on the rear panel of new AXP pump module.
- Connect No. 9(Ground) in 25-pin D-Sub Port to AXP No. 6 pin port(Common Replay).
- Connect No. 23(Bottom Pump Stop) in 25-pin D-Sub Port to AXP No.8 pin(Normally open).
- Connect the AXP pump, ICS-6000 pump in Chromeleon. Plug in the 25-pin D-Sub Port in ICS-6000 pump rear panel and connect the other 10-pin end to AXP rear panel.
- AXP pump will stop its motor when its pressure is higher than HighPressureLimit or after 20 strokes when the pressure is lower than LowPressureLimit.
- A trigger signal will be sent out automatically from AXP to ICS-6000 pump when the AXP pump motor is stopped by a high pressure error or low pressure error.
- The trigger signal will stop ICS-6000 BOTTOM pump motor.
- To stop the ICS-6000 TOP pump motor in a DP module, replace No.23(Bottom Pump Stop) pin connection by No. 24(Top Pump Stop) pin in step 6.

