Skip to main content
Thermo Fisher Scientific

Troubleshoot Air in the Dionex ICS-6000 Pump or Fluid Path Causing Unstable Pressure or an Oscillating Conductivity Baseline

Use this article when a Dionex ICS-6000 has visible or suspected air in the fluid path and the pump pressure and/or conductivity baseline is unstable, pulsating, noisy, or oscillating. The troubleshooting process helps determine whether air is entering upstream of the pump, the pump is losing prime or being starved, or gas is trapped downstream in the suppressor/conductivity-cell path. The key diagnostic is to compare the pump pressure trace and conductivity trace at the same time, then isolate

Issue

“My Dionex ICS-6000 has air or bubbles in the fluid path and the baseline is unstable or oscillating.”

“The conductivity baseline is pulsating or noisy, and I am not sure whether the air is coming from the pump or is trapped in the detector cell.”

“The pump pressure may also be fluctuating, or the system improves after purging but the air or baseline instability comes back.”

“I need to determine where the air is entering the system and whether it is affecting pump delivery or only the conductivity detector.”

Environment

  • Thermo Scientific Dionex ICS-6000
  • ICS-6000 SP or ICS-6000 DP
  • Conductivity detection
  • Suppressed conductivity applications, when applicable
  • RFIC/eluent generation, when installed
  • Chromeleon CDS 7.x, including environments such as Chromeleon 7.3.2
  • AS-AP, AS-DV, manual injection, or other sample introduction may be installed, but the sampler is usually secondary unless the symptom occurs only during injection

Resolution

  • Determine whether the baseline behavior is abnormal for the method.

    • Confirm whether the method is isocratic or gradient.
    • For an isocratic method, allow the system to equilibrate and verify whether the conductivity baseline continues to oscillate, pulse, or drift.
    • For a gradient method, distinguish normal programmed baseline movement from repetitive or irregular oscillation.
    • If possible, observe the baseline without injections so that the autosampler and sample matrix are removed from the first troubleshooting step.
  • Monitor pump pressure and conductivity at the same time in Chromeleon.

    • Display or record the pump pressure trace.
    • Display or record the conductivity trace over the same time period.
    • If applicable, also monitor suppressor current/state, EGC/CR-TC state, temperature, and flow command.
    • Compare the timing of the conductivity disturbance with the pressure fluctuation.

    Interpret the result as follows:

    • Pressure and conductivity fluctuate together: prioritize an upstream hydraulic problem such as air entering the pump, incomplete priming, inlet starvation, inadequate degassing, or pump-delivery instability.
    • Pressure is stable but conductivity oscillates: prioritize trapped gas in the conductivity-cell/suppressor path, downstream outgassing, detector flow-path issues, or suppression-related problems.
  • Inspect the low-pressure eluent path from the reservoir to the pump.

    • Verify the eluent reservoir contains sufficient liquid.
    • Confirm the pickup tubing remains submerged.
    • Inspect the end-line filter for contamination or restriction.
    • Check tubing for kinks, cracks, loose fittings, or damaged connections.
    • Watch the tubing while the pump is running.
    • Identify the first point at which a bubble appears rather than only observing where the bubble eventually travels.
    • Verify the eluent is properly prepared and degassed.
    • Verify the installed degasser is operating correctly for the specific ICS-6000 pump configuration.
  • Purge and fully prime the pump using the applicable ICS-6000 procedure.

    • Use properly prepared and degassed eluent.
    • Open the appropriate purge path.
    • Purge until visible bubbles are removed.
    • Close the purge path.
    • Return the pump to the normal method flow rate.
    • Observe pressure and conductivity again.

    Interpret the result:

    • If pressure and baseline become stable and remain stable, trapped air or incomplete prime was likely involved.
    • If the system improves after purge but bubbles or instability return, do not continue repeated purging as the main corrective action. The system is likely reintroducing air or starving the pump. Return to the reservoir, low-pressure tubing, fittings, end-line filter, and degasser path.
    • If pressure remains unstable after air has been removed, continue with pump hydraulic isolation.
  • Measure actual pump flow if pump delivery is still in question.

    • Use a safe, documented fluidic boundary downstream of the pump.
    • Collect effluent for a known period of time.
    • Compare measured flow with the commanded flow.
    • Repeat if necessary to determine whether actual flow is stable.

    Interpretation:

    • Actual flow unstable: continue in the pump/hydraulic branch. Possible remaining causes include persistent air, inlet starvation, check-valve behavior, piston-seal leakage, or pump-head hydraulic problems.
    • Actual flow stable but pressure is unusually unstable: investigate pressure-sensing or pulse-damping behavior before assuming a gross flow-delivery failure.

    Do not change pump calibration to compensate for unresolved hydraulic instability.

  • If pump pressure and flow are stable, evaluate trapped air in the conductivity cell.

    • Inspect tubing leading into and out of the conductivity detector.
    • Look for bubbles entering or leaving the cell.
    • Confirm the flow path is plumbed correctly.
    • Verify the downstream tubing/backpressure arrangement is appropriate for the installed suppressor and application.
    • Follow the applicable ICS-6000 conductivity-cell procedure for removing trapped air.

    Trapped air in the conductivity cell can produce:

    • baseline pulsation
    • random noise
    • unstable conductivity readings
    • low or inconsistent detector response

    If the pressure remains stable and removing the trapped cell air restores a stable conductivity baseline, the pump was not the primary cause.

  • Evaluate suppressor and downstream backpressure conditions.

    • Confirm the exact suppressor model and generation.
    • Check the suppressor outlet and conductivity-cell outlet tubing for restriction.
    • Verify that downstream backpressure is neither excessive nor too low for the specific suppressor/application.
    • Use the exact suppressor documentation for numerical pressure limits.

    Insufficient downstream pressure can allow dissolved gas to come out of solution in some configurations, while excessive downstream restriction can create a different fluidic problem or stress the suppressor.

  • If air has been excluded and pump delivery is still unstable, continue with pump hydraulic troubleshooting.
    Evaluate the remaining causes one at a time, as appropriate for the installed ICS-6000 pump:

    • inlet check valve
    • outlet check valve
    • piston seals
    • piston condition
    • pump-head leakage
    • pulse-damping component
    • pressure-sensing behavior
    • pump mechanical operation

    Do not replace multiple pump components at the same time when one isolation test can determine which branch remains likely.

  • Verify the correction under the original method.

    • Restore the original column, suppressor, detector, and method configuration.
    • Run the normal flow rate.
    • Verify pressure is stable.
    • Verify actual flow if it was previously abnormal.
    • Confirm visible air is no longer present.
    • Confirm the conductivity baseline is stable.
    • Run representative standards.
    • Verify retention-time repeatability, peak area repeatability, peak shape, conductivity background, and pressure behavior.
    • For gradient methods, compare the full gradient baseline profile run-to-run rather than expecting a flat baseline.

Analysis

Cause