Troubleshoot High or Noisy Background or Poor Response After Installing or Servicing an NGES Suppressor on an ICS-6000

Use this article when a Dionex Next Generation Electrolytic Suppressor (NGES) has been installed, replaced, reconnected, or returned to service on an ICS-6000 and the system shows high background conductivity, noisy or drifting baseline, low/increasing peak areas, or poor sensitivity. The NGES requires a specific hydration and backpressure-check procedure during initial installation and after an extended idle period of 7 days or more. This article covers the required startup procedure and the tr

Issue

“My Dionex ICS-6000 has air or bubbles in the fluid path and the baseline is unstable or oscillating.”

“The conductivity baseline is pulsating or noisy, and I am not sure whether the air is coming from the pump or is trapped in the detector cell.”

“The pump pressure may also be fluctuating, or the system improves after purging but the air or baseline instability comes back.”

“I need to determine where the air is entering the system and whether it is affecting pump delivery or only the conductivity detector.”

Environment

  • Thermo Scientific Dionex ICS-6000
  • Dionex NGES Next Generation Electrolytic Suppressor
  • NGES-A for applicable anion applications
  • NGES-C for applicable cation applications
  • 2 mm or 4 mm NGES configuration
  • AutoSuppression recycle mode or AutoSuppression external water mode
  • Chromeleon 7.2.10 MUj
  • Chromeleon 7.3.2 MUe

Resolution

Note: On an ICS-6000 with an earlier supported Chromeleon version, the NGES can still be used but defaults to ERS identification rather than being recognized specifically as NGES.

  • Determine whether the NGES requires the startup/hydration procedure.

    • Perform the NGES startup procedure:
      • during initial installation, or
      • after an extended idle period of 7 days or more.
    • Do not assume that simply connecting the suppressor and flowing eluent is sufficient.
    • The hydration procedure is required to place the NGES ion-exchange materials into the properly swollen state for operation. Improper hydration can damage the suppressor. Thermo Fisher
  • Configure the NGES temporarily for hydration.
    Make the following flow connections:

    • Analytical column OUTLET → NGES ELUENT IN
    • NGES ELUENT OUT → NGES REGEN IN
    • NGES REGEN OUT → waste

    Turn the NGES power OFF before beginning hydration.

    Important: Do not power the NGES during the hydration step. The NGES Operating Manual warns that operating it electrically during hydration can cause irreversible damage. Thermo Fisher

  • Hydrate the NGES.
    Run either DI water or eluent at 10 mM or less through the suppressor for 20 minutes at:

    • 4 mm NGES: 0.5 mL/min
    • 2 mm NGES: 0.125 mL/min

    Then:

    • Stop the pump.
    • Allow the NGES to sit undisturbed for 20 minutes. Thermo Fisher
  • Measure the NGES backpressure before returning it to operation.

    • Start the pump.
    • Run at the application flow rate for approximately 2 minutes.
    • Record the required pressure readings using the NGES startup/backpressure procedure.
    • Calculate the pressure drop across the NGES.

    The NGES pressure drop should be ≤200 psi. If it is higher, do not continue normal operation until the excessive backpressure has been investigated. Thermo Fisher

  • Restore the normal AutoSuppression flow path.
    For AutoSuppression recycle mode:

    • NGES ELUENT OUT → conductivity detector INLET
    • Conductivity detector OUTLET → required backpressure coil(s) → NGES REGEN IN
    • NGES REGEN OUT → downstream regeneration path/gas separator waste configuration as specified for the system.

    Ensure the electrolytic gas separator waste tube is installed correctly so that generated gases are safely vented. Thermo Fisher

    If using AutoSuppression external water mode, restore the applicable external-water plumbing before operation.

  • Verify Chromeleon configuration after installing a new NGES.
    For an ICS-6000:

    • Open Chromeleon Configuration Manager.
    • Select the suppressor configuration.
    • Open the Device Configuration window.
    • Confirm the suppressor configuration.
    • Save the configuration changes.

    Use an NGES-compatible Instrument Method and verify that the suppressor is operating in constant-current mode with the appropriate current for the application. Thermo Fisher

  • Verify that the suppressor current is correct for the eluent and flow rate.

    • Confirm the eluent chemistry.
    • Confirm eluent concentration.
    • Confirm method flow rate.
    • Compare the programmed current with the NGES current requirement in the current NGES Operating Manual.
    • For gradients, ensure the selected current is appropriate for the maximum eluent concentration encountered in the gradient.

    Incorrect or insufficient suppressor current can cause:

    • high background
    • incomplete suppression
    • reduced response
    • progressively increasing peak areas

    The NGES Operating Manual specifically identifies incorrect current as a cause of high background conductivity. Thermo Fisher

  • Verify that eluent does not flow through the NGES for an extended period with suppressor power off.
    During normal AutoSuppression operation:

    • Start eluent flow and turn on NGES power immediately.
    • For external-water mode, establish regenerant-water flow first, then start eluent flow and immediately turn on suppressor power.

    Do not allow eluent to flow through the NGES unpowered for more than a few minutes. Extended operation with eluent flowing while the suppressor is off can consume suppressor capacity and produce low or increasing analyte peak areas. Thermo Fisher

  • If background remains high, isolate suppression from general system problems.
    Verify:

    • suppressor power is on
    • correct current is applied
    • eluent is entering NGES ELUENT IN
    • eluent exits NGES ELUENT OUT
    • there are no leaks
    • eluent chemistry and purity are correct
    • tubing is connected to the correct ports
    • conductivity cell and outlet path are not excessively restricted

    The NGES manual identifies the following high-background causes:

    • incorrect current
    • flow-path blockage
    • broken suppressor seal
    • contaminated membrane
    • poor eluent quality. Thermo Fisher
  • Check downstream backpressure.
    The conductivity detector, tubing, and backpressure coils connected to NGES ELUENT OUT all contribute to downstream pressure.

    Verify that:

    • backpressure coils are correctly configured
    • detector cell is not restricted
    • outlet tubing is not blocked
    • NGES pressure drop remains ≤200 psi

    For the NGES backpressure-coil portion of the flow path, the current manual specifies adjustment when the coil pressure is:

    • >40 psi: shorten the backpressure tubing
    • <30 psi with carbonate eluent: add tubing
    • <10 psi with other eluents: add tubing. Thermo Fisher

    Do not confuse the backpressure-coil requirement with the total allowable NGES pressure drop.

  • If the conductivity detector appears restricted, isolate the detector cell.
    The NGES troubleshooting procedure permits detector-cell backflushing where applicable:

    • Analytical column OUTLET → conductivity detector OUTLET
    • Conductivity detector INLET → waste
    • Run the application flow for approximately 10 minutes.
    • Re-measure detector-cell backpressure.

    If detector-cell backpressure remains >70 psi after repeated supported backflush attempts, the NGES manual directs replacement of the detector cell. Thermo Fisher

  • If NGES backpressure itself remains high, backflush the suppressor with DI water.
    Using an external pump:

    • External pump → NGES ELUENT OUT
    • NGES ELUENT IN → waste

    Run DI water:

    • 4 mm: 1.0 mL/min for 10 minutes
    • 2 mm: 0.25 mL/min for 10 minutes

    Re-measure NGES pressure drop.

    Repeat if necessary. If NGES pressure remains >200 psi after repeated supported backflushing, replace the NGES. Thermo Fisher

  • If the problem is low or progressively increasing response, regenerate the NGES.
    Loss of suppressor capacity can occur if:

    • eluent flowed through the NGES while power was off, or
    • insufficient current was applied for an extended period.

    Start with electrolytic regeneration when appropriate:

    • Run the system under normal application conditions.
    • For a gradient method, set the eluent concentration to the maximum value normally used in the gradient.
    • Increase suppressor current to 60% above the optimum current.
    • Run for 3–6 hours.
    • Return the current to the normal optimum value. Thermo Fisher

    Do not exceed the NGES maximum current:

    • NGES-A 4 mm: 500 mA
    • NGES-C 4 mm: 300 mA
    • NGES-A 2 mm: 150 mA
    • NGES-C 2 mm: 110 mA. Thermo Fisher
  • If routine electrolytic regeneration does not restore performance, escalate to the supported regeneration procedure.
    The NGES manual provides:

    • chemical regeneration
    • electrolytic regeneration
    • full/extended regeneration.

    Full/extended regeneration is intended for greater capacity loss when electrolytic regeneration is insufficient. Use the exact NGES-A or NGES-C procedure and the required chemical for the suppressor type. Do not improvise regeneration chemistry. Thermo Fisher

  • If overall sensitivity remains low, evaluate the complete system before condemning the NGES.
    Check:

    • system leaks
    • injection-valve operation
    • suppressor regeneration status
    • backpressure-coil configuration
    • suppressor membrane condition.

    The NGES manual specifically recommends regeneration and checking that backpressure coils are not exceeding 40 psi when troubleshooting decreased sensitivity for all analytes. If sensitivity remains low after the prescribed checks and cleaning, replace the NGES. Thermo Fisher

  • Verify the correction with the original application.

    • Return current to the proper application setting.
    • Allow the system to equilibrate.
    • Monitor pressure and conductivity.
    • Confirm that background returns to the normal range for the method.
    • Run at least two injections of an appropriate standard.
    • Compare:
      • peak area
      • peak height
      • peak width
      • retention time
      • background conductivity
      • baseline noise

    If the correct response does not return after the supported regeneration procedure and two test-standard injections, escalate according to the current NGES Operating Manual.

Analysis

Cause